When to Perform Control Solution Testing:
Performing a Control Solution Test:
Step 1: Preparation
Step 2: Insert Test Strip
Step 3: Set Monitor
Step 4: Apply Control Solution
Step 5: Compare Results
Glucose Control Solution Troubleshooting:
Possible Causes | What you can do |
---|---|
Wrong brand of Control Solution being used | Make sure you are using Embrace TALK Control Solution. |
Control Solution not at room temperature | Make sure the testing environment is between 50�F to 104�F, <85% RH. |
Expired or contaminated control solution | Check the expiration and open bottle date on both the control solution and test strips. (Test strips should be used within 6 months from date opened. Control solution should be used within 3 months from date opened). Repeat testing using a new test strip. If the result is still out of range, use a new bottle of control solution and retest. |
Damaged test strip | |
Monitor malfunction | If problem persists, call Omnis Health@1-877-979-5454. |